.. _CreateIkedaCarpenterParameters: Create Ikeda Carpenter Parameters ================================= .. contents:: Table of Contents :local: Example of how a parameter can be determined on a beamline ---------------------------------------------------------- The steps involved in creating instrument specific parameters for the :ref:`Ikeda-Carpenter-Pseudo-Voigt ` peak shape function are on ISIS/GEM approximately as follows: - Run calibration samples: NIST standard Silicon and Y2O3. - Combined rietveld refined of these two standards in either Fullprof or GSAS - All structural parameters are fixed but not the Ikeda Carpenter parameters and background parameters (and zero shifts). On GEM, the Ikeda-Carpenter parameters are the same for all banks but the Pseudo-Voigt sigma and gamma parameters which are refined differently for the different banks. - Finally the information from this refinement is copied and pasted from output files to generate either a :ref:`Fullprof ` or :ref:`GSAS ` instrument parameter files. .. categories:: Techniques