Calibration

Calibration is currently based on tubes and is accessed using the tube module.

Definition of Calibration

tube.calibrate(ws, tubeSet, knownPositions, funcForm[, fitPar, margin, rangeList, calibTable, plotTube, excludeShorTubes, overridePeaks, fitPolyn, outputPeak])

Define the calibrated positions of the detectors inside the tubes defined in tubeSet.

Tubes may be considered a list of detectors aligned that may be considered as pixels for the analogy when they values are displayed.

The position of these pixels are provided by the manufacturer, but its real position depends on the electronics inside the tube and varies slightly from tube to tube. The calibrate method, aims to find the real positions of the detectors (pixels) inside the tube.

For this, it will receive an Integrated workspace, where a special measurement was performed so to have a pattern of peaks or through. Where gaussian peaks or edges can be found.

The calibration follows the following steps

  1. Finding the peaks on each tube
  2. Fitting the peaks against the Known Positions
  3. Defining the new position for the pixels(detectors)

Let’s consider the simplest way of calling calibrate:

from tube import calibrate
ws = Load('WISH17701')
ws = Integration(ws)
known_pos = [-0.41,-0.31,-0.21,-0.11,-0.02, 0.09, 0.18, 0.28, 0.39 ]
peaks_form = 9*[1] # all the peaks are gaussian peaks
calibTable = calibrate(ws,'WISH/panel03',known_pos, peaks_form)

In this example, the calibrate framework will consider all the tubes (152) from WISH/panel03. You may decide to look for a subset of the tubes, by passing the rangeList option.

# This code will calibrate only the tube indexed as number 3
# (usually tube0004)
calibTable = calibrate(ws,'WISH/panel03',known_pos,
                       peaks_form, rangeList=[3])

Finding the peaks on each tube

Dynamically fitting peaks

The framework expects that for each tube, it will find a peak pattern around the pixels corresponding to the known_pos positions.

The way it will work out the estimated peak position (in pixel) is:

  1. Get the length of the tube: distance(first_detector,last_detector) in the tube.
  2. Get the number of detectors in the tube (nDets)
  3. It will be assumed that the center of the tube correspond to the origin (0)
centre_pixel = known_pos * nDets/tube_length + nDets/2

It will them look for the real peak around the estimated value as:

# consider tube_values the array of counts, and peak the estimated
# position for the peak
real_peak_pos = argmax(tube_values[peak-margin:peak+margin])

After finding the real_peak_pos, it will try to fit the region around the peak to find the best expected position of the peak in a continuous space. It will do this by fitting the region around the peak to a Gaussian Function, and them extract the PeakCentre returned by the Fitting.

centre = real_peak_pos
fit_start, fit_stop = centre-margin, centre+margin
values = tube_values[fit_start,fit_stop]
background = min(values)
peak = max(values) - background
width = len(where(values > peak/2+background))
# It will fit to something like:
# Fit(function=LinerBackground,A0=background;Gaussian,
# Height=peak, PeakCentre=centre, Sigma=width,fit_start,fit_end)

Force Fitting Parameters

These dynamically values can be avoided by defining the fitPar for the calibrate function

eP = [57.5, 107.0, 156.5, 206.0, 255.5, 305.0, 354.5, 404.0, 453.5]
# Expected Height of Gaussian Peaks (initial value of fit parameter)
ExpectedHeight = 1000.0
# Expected width of Gaussian peaks in pixels
# (initial value of fit parameter)
ExpectedWidth = 10.0
fitPar = TubeCalibFitParams( eP, ExpectedHeight, ExpectedWidth )
calibTable = calibrate(ws, 'WISH/panel03', known_pos, peaks_form, fitPar=fitPar)

Different Function Factors

Although the examples consider only Gaussian peaks, it is possible to change the function factors to edges by passing the index of the known_position through the funcForm. Hence, considering three special points, where there are one gaussian peak and two edges, the calibrate could be configured as

known_pos = [-0.1 2 2.3]
# gaussian peak followed by two edges (through)
form_factor = [1 2 2]
calibTable = calibrate(ws,'WISH/panel03',known_pos, form_factor)

Override Peaks

It is possible to scape the finding peaks position steps by providing the peaks through the overridePeaks parameters. The example below tests the calibration of a single tube (30) but skips the finding peaks step.

known_pos = [-0.41,-0.31,-0.21,-0.11,-0.02, 0.09, 0.18, 0.28, 0.39 ]
define_peaks = [57.5, 107.0, 156.5, 206.0, 255.5, 305.0, 354.5,
               404.0, 453.5]
calibTable = calibrate(ws, 'WISH/panel03', known_pos, peaks_form,
                 overridePeaks={30:define_peaks}, rangeList=[30])

Output Peaks Positions

Enabling the option outputPeak a WorkspaceTable will be produced with the first column as tube name and the following columns with the position where corresponding peaks were found. Like the table below.

TubeId Peak1 ... PeakM
tube0 15.5 ... 370.3
... ... ... ...
tubeN 14.9 ... 371.2

The signature changes to:

.. code-block:: python
calibTable, peakTable = calibrate(...)

It is possible to give a peakTable directly to the outputPeak option, which will make the calibration to append the peaks to the given table.

Hint

It is possible to save the peakTable to a file using the savePeak() method.

Find the correct position along the tube

The second step of the calibration is to define the correct position of pixels along the tube. This is done by fitting the peaks positions found at the previous step against the known_positions provided.

known       |              *
positions   |           *
            |      *
            |  *
            |________________
              pixels positions

The default operation is to fit the pixels positions against the known positions with a quadratic function in order to define an operation to move all the pixels to their real positions. If necessary, the user may select to fit using a polinomial of 3rd order, through the parameter fitPolyn.

Note

The known positions are given in the same unit as the spacial position (3D) and having the center of the tube as the origin.

Hence, this section will define a function that:

F(pix) = RealRelativePosition

Define the new position for the detectors

Finally, the position of the detectors are defined as a vector operation like

\vec{p} = \vec{c} + v \vec{u}

Where \vec{p} is the position in the 3D space, v is the RealRelativePosition deduced from the last session, and finally, \vec{u} is the unitary vector in the direction of the tube.

Parameters:
  • ws – Integrated workspace with tubes to be calibrated.
  • tubeSet – Specification of Set of tubes to be calibrated. If a string is passed, a TubeSpec will be created passing the string as the setTubeSpecByString.

This will be the case for TubeSpec as string

self.tube_spec = TubeSpec(ws)
self.tube_spec.setTubeSpecByString(tubeSet)

If a list of strings is passed, the TubeSpec will be created with this list:

self.tube_spec = TubeSpec(ws)
self.tube_spec.setTubeSpecByStringArray(tubeSet)

If a TubeSpec object is passed, it will be used as it is.

Parameters:
  • knownPositions – The defined position for the peaks/edges, taking the center as the origin and having the same units as the tube length in the 3D space.
  • funcForm – list with special values to define the format of the peaks/edge (peaks=1, edge=2). If it is not provided, it will be assumed that all the knownPositions are peaks.

Optionals parameters to tune the calibration:

Parameters:
  • fitPar – Define the parameters to be used in the fit as a TubeCalibFitParams. If not provided, the dynamic mode is used. See provideTheExpectedValue()
  • margin – value in pixesl that will be used around the peaks/edges to fit them. Default = 15. See the code of TubeCalibDemoMerlin where margin is used to calibrate small tubes.
fit_start, fit_end = centre - margin, centre + margin
Parameters:rangeList – list of tubes indexes that will be calibrated. As in the following code (see: improvingCalibrationSingleTube()):
for index in rangelist:
    do_calibrate(tubeSet.getTube(index))
Parameters:
  • calibTable – Pass the calibration table, it will them append the values to the provided one and return it. (see: TubeCalibDemoMerlin)
  • plotTube – If given, the tube whose index is in plotTube will be ploted as well as its fitted peaks, it can receive a list of indexes to plot.(see: changeMarginAndExpectedValue())
  • excludeShortTubes – Do not calibrate tubes whose length is smaller than given value. (see at Examples/TubeCalibDemoMerlin_Adjustable.py)
  • overridePeaks – dictionary that defines an array of peaks positions (in pixels) to be used for the specific tube(key). (see: improvingCalibrationSingleTube())
for index in rangelist:
  if overridePeaks.has_key(index):
    use_this_peaks = overridePeaks[index]
    # skip finding peaks
    fit_peaks_to_position()
Parameters:
  • fitPolyn – Define the order of the polynomial to fit the pixels positions against the known positions. The acceptable values are 1, 2 or 3. Default = 2.
  • outputPeak – Enable the calibrate to output the peak table, relating the tubes with the pixels positions. It may be passed as a boolean value (outputPeak=True) or as a peakTable value. The later case is to inform calibrate to append the new values to the given peakTable. This is useful when you have to operate in subsets of tubes. (see TubeCalibDemoMerlin that shows a nice inspection on this table).
calibTable, peakTable = calibrate(ws, (omitted), rangeList=[1],
         outputPeak=True)
# appending the result to peakTable
calibTable, peakTable = calibrate(ws, (omitted), rangeList=[2],
         outputPeak=peakTable)
# now, peakTable has information for tube[1] and tube[2]
Return type:calibrationTable, a TableWorkspace with two columns DetectorID(int) and DetectorPositions(V3D).

Use Cases

Among the examples, inside the Examples folder, the user is encouraged to look at TubeCalibDemoMaps_All, there he will find 7 examples showing how to use calibrate method.

  • minimalInput() shows the easiest way to use calibrate.
  • provideTheExpectedValue() shows the usage of fitPar parameter to provide the expected values for the peaks in pixels.
  • changeMarginAndExpectedValue() demonstrate how to use margin, fitPar, plotTube, and outputPeak
  • improvingCalibrationSingleTube() explores the usage of rangeList and overridePeaks to improve the calibration of specific tubes.
  • improvingCalibrationOfListOfTubes() extends theimprovingCalibrationSingleTube to provide a good calibration to almost all instrument.
  • calibrateB2Window() explore a singularity of the MAP14919 example, where the second peak does not appear clear on some tubes inside one door. So, this example, shows how to use rangeList to carry a calibration to the group of tubes.
  • completeCalibration() demonstrate how the rangeList, overridePeaks, may be used together to allow the calibration of the whole instrument, despite, its particularities in some cases.
  • findThoseTubesThatNeedSpecialCareForCalibration() show an approach to find the tubes that will require special care on calibrating. It will also help to find detectors that are not working well.

Examples

MAPS instrument

Tube Calibration Demonstration for MAPS instrument.

This module group many examples and also demonstrate how to work with the tube.calibrate().

It starts from the simplest way to perform a calibration. It them increase the complexity of dealing with real calibration work, when particularities of the instrument must be considered

At the end, it gives a suggestion on a techinique to investigate and improve the calibration itself.

Minimal Input

Examples.TubeCalibDemoMaps_All.minimalInput(filename)

Simplest way of calling tube.calibrate()

The minimal input for the calibration is the integrated workspace and the known positions.

Even though it is easy to call, the calibration performs well, but there are ways to improve the results, as it is explored after.

../_images/outputOfMinimalInput.png

Provide Expected Value

Examples.TubeCalibDemoMaps_All.provideTheExpectedValue(filename)

Giving the expected value for the position of the peaks in pixel.

The minimalInput() let to the calibrate to guess the position of the pixels among the tubes. Although it works nicelly, providing these expected values may improve the results. This is done through the fitPar parameter.

Change Margin and Expected Value

Examples.TubeCalibDemoMaps_All.changeMarginAndExpectedValue(filename)

To fit correctly, it is important to have a good window around the peak. This windown is defined by the margin parameter.

This examples shows how the results worsen if we change the margin from its default value 15 to 10.

It shows how to see the fitted values using the plotTube parameter.

It will also output the peaks position and save them, through the outputPeak option and the tube.savePeak() method.

An example of the fitted data compared to the acquired data to find the peaks positions:

../_images/calibratePlotFittedData.png

The result deteriorate, as you can see:

../_images/calibrateChangeMarginAndExpectedValue.png

Improving Calibration Single Tube

Examples.TubeCalibDemoMaps_All.improvingCalibrationSingleTube(filename)

The provideTheExpectedValue() provided a good solution, but there are few tubes whose calibration was not so good.

This method explores how to deal with these tubes.

First of all, it is important to identify the tubes that did not work well.

From the outputs of provideTheExpectedValue, looking inside the instrument tree, it is possible to list all the tubes that are not so good.

Unfortunately, they do not have a single name identifier. So, locating them it is a little bit trickier. The findThoseTubesThatNeedSpecialCareForCalibration() shows one way of finding those tubes. The index is the position inside the PeakTable.

For this example, we have used inspection from the Instrument View. One of them is inside the A1_Window, 3rd PSD_TUBE_STRIP 8 pack up, 4th PSD_TUBE_STRIP: Index = 8+8+4 - 1 = 19.

In this example, we will ask the calibration to run the calibration only for 3 tubes (indexes 18,19,20). Them, we will check why the 19 is not working well. Finally, we will try to provide another peaks position for this tube, and run the calibration again for these tubes, to improve the results.

This example shows how to use overridePeaks option

Improving Calibration of List of Tubes

Examples.TubeCalibDemoMaps_All.improvingCalibrationOfListOfTubes(filename)

Analysing the result of provideTheExpectedValue it was seen that the calibration of some tubes was not good.

Note

This method list some of them, there are a group belonging to window B2 that shows only 2 peaks that are not dealt with here.

If first plot the bad ones using the plotTube option. It them, find where they fail, and how to correct their peaks, using the overridePeaks. If finally, applies the calibration again with the points corrected.

Calibrating Window B2

Examples.TubeCalibDemoMaps_All.calibrateB2Window(filename)

There are among the B2 window tubes, some tubes that are showing only 2 strips.

Those tubes must be calibrated separated, as the known positions are not valid.

This example calibrate them, using only 4 known values: 2 edges and 2 peaks.

Run this example, and them see the worksapce in the calibrated instrument and you will see how it worked.

The picture shows the output, look that only a section of the B2 Window was calibrated.

../_images/calibrateB2Window.png

Complete Calibration

Examples.TubeCalibDemoMaps_All.completeCalibration(filename)

This example shows how to use some properties of calibrate method to join together the calibration done in provideTheExpectedValue(), and improved in calibrateB2Window(), and improvingCalibrationOfListOfTubes().

It also improves the result of the calibration because it deals with the E door. The acquired data cannot be used to calibrate the E door, and trying to do so, produces a bad result. In this example, the tubes inside the E door are excluded to the calibration. Using the ‘’‘rangeList’‘’ option.

Calibration technique: Finding tubes not well calibrated

Examples.TubeCalibDemoMaps_All.findThoseTubesThatNeedSpecialCareForCalibration(filename)

The example provideTheExpectedValue() has shown its capability to calibrate almost all tubes, but, as explored in the improvingCalibrationOfListOfTubes() and improvingCalibrationSingleTube() there are some tubes that could not be calibrated using that method.

The goal of this method is to show one way to find the tubes that will require special care.

It will first perform the same calibration seen in provideTheExpectedValue(), them, it will process the peakTable output of the calibrate method when enabling the parameter outputPeak.

It them creates the Peaks workspace, that is the difference of the peaks position from the expected values of the peaks positions for all the tubes. This allows to spot what are the tubes whose fitting are outliers in relation to the others.

../_images/plotingPeaksDifference.png

The final result for this method is to output using plotTube the result of the fitting to all the ‘outliers’ tubes.

MERLIN instrument

Tube Calibration Demonstration program for MERLIN.

Attention

MERLIN instruments are loaded with already calibrated values. The calibration works nicelly with these files, but if you want to see the uncalibrated file you can do it. Look at How to reset detectors calibration.

In this example, the calibration of the whole MERLIN instrument is shown. It demonstrate how to use tube.calibrate() to calibrate MERLIN tubes.

Opening the calibrated data in the Instrument View, it is possible to group the tubes in some common regions:

  • Doors 9 and 8 are similar and can be calibrated together using 7 key points
  • Doors 7,6,5,4, 2 and 1 can be calibrated using 9 key points.
  • Door 3 is particular, because it is formed with some smaller tubes as well as some large tubes.

This example shows:

  • How to calibrate regions of the instrument separetelly.

  • How to use calibTable parameter to append information in order to create a calibration table for the whole instrument.

  • How to use the outputPeak to check how the calibration is working as well as the usage of analisePeakTable method to

    look into the details of the operation to improve the calibration.

  • It deals with defining different known positions for the different tube lengths.

The output of this examples shows an improvement in relation to the previous calibrated instrument.

../_images/calibratedMantidMerlin1.png

The previous calibrated instrument view:

../_images/calibratedCurrentMerlin.png

Other Useful Methods

tube.savePeak(peakTable, filePath)

Allows to save the peakTable to a text file.

Parameters:peakTable – peak table as the workspace table provided by calibrated method, as in the example:
calibTable, peakTable = calibrate(..., outputPeak=peakTable)
savePeak(peakTable, 'myfolder/myfile.txt')
Parameters:filePath – where to save the file. If the filePath is not given as an absolute path, it will be considered relative to the defaultsave.directory.

The file will be saved with the following format:

id_name (parsed space to %20) [peak1, peak2, ..., peakN]

You may load these peaks using readPeakFile

panel1/tube001 [23.4, 212.5, 0.1]
...
panel1/tubeN   [56.3, 87.5, 0.1]
tube.readPeakFile(file_name)

Load the file calibration

It returns a list of tuples, where the first value is the detector identification and the second value is its calibration values.

Example of usage:

for (det_code, cal_values) in readPeakFile('pathname/TubeDemo'):
   print(det_code)
   print(cal_values)
Parameters:file_name – Path for the file
Return type:list of tuples(det_code, peaks_values)
tube.saveCalibration(table_name, out_path)

Save the calibration table to file

This creates a CSV file for the calibration TableWorkspace. The first column is the detector number and the second column is the detector position

Example of usage:

saveCalibration('CalibTable','/tmp/myCalibTable.txt')
Parameters:
  • table_name – name of the TableWorkspace to save
  • out_path – location to save the file
tube.readCalibrationFile(table_name, in_path)

Read a calibration table from file

This loads a calibration TableWorkspace from a CSV file.

Example of usage:

saveCalibration('CalibTable','/tmp/myCalibTable.txt')
Parameters:
  • table_name – name to call the TableWorkspace
  • in_path – the path to the calibration file

Implementation

More details on the finer points of the calibration implementation can be found at http://www.mantidproject.org/Tube_Calibration and in:

Category: Techniques