Name | Direction | Type | Default | Description |
---|---|---|---|---|
InputWorkspace | Input | MatrixWorkspace | Mandatory | |
OutputWorkspace | Output | MatrixWorkspace | Mandatory | |
DetectorTubes | Input | boolean | False | If true, the algorithm will assume that the detectors are tubes in the Y direction. |
DetectorWing | Input | boolean | False | If true, the algorithm will assume that the detector is curved around the sample. E.g. BIOSANS Wing detector. |
OutputMessage | Output | string | ||
ReductionProperties | Input | string | __sans_reduction_properties |
Performs a solid angle correction on all detector pixels. This algorithm is usually called by SANSReduction or HFIRSANSReduction.
The solid angle correction is applied as follows:
\(I'(x,y)=\frac{I(x,y)}{\cos^3(2\theta)}\)
\(\sigma_{I'(x,y)}=\frac{\sigma_{I(x,y)}}{\vert\cos^3(2\theta)\vert}\)
If DetectorTubes is set to True, the correction is calculated according to a tube geometry. The cosine term above then becomes:
\(\cos^3(2\theta) \rightarrow \cos^2(2\theta) \cos(\alpha)\)
where \(\alpha\): is the angle between the sample-to-pixel vector and its projection on the X-Z plane.
Categories: AlgorithmIndex | Workflow\SANS\UsesPropertyManager | CorrectionFunctions\InstrumentCorrections