\(\renewcommand\AA{\unicode{x212B}}\)

HB3AIntegratePeaks v1

../_images/HB3AIntegratePeaks-v1_dlg.png

HB3AIntegratePeaks dialog.

Summary

Integrates peaks from the input MDEvent workspace and can optionally apply a Lorentz correction to the output peaks workspace; output can be saved in different formats.

See Also

HB3AFindPeaks, IntegratePeaksMD, HB3AAdjustSampleNorm, HB3APredictPeaks

Properties

Name

Direction

Type

Default

Description

InputWorkspace

Input

str list

Mandatory

Input MDEvent workspace to use for integration

PeaksWorkspace

Input

str list

Mandatory

Peaks workspace containing peaks to integrate

PeakRadius

Input

number

1

Fixed radius around each peak position in which to integrate (same units as input workspace)

BackgroundInnerRadius

Input

number

0

Inner radius used to evaluate the peak background

BackgroundOuterRadius

Input

number

0

Outer radius used to evaluate the peak background

ApplyLorentz

Input

boolean

True

Whether the Lorentz correction should be applied to the integrated peaks

RemoveZeroIntensity

Input

boolean

True

If to remove peaks with 0 or less intensity from the output

OutputFormat

Input

string

SHELX

Save direction cosines in HKL, or the fullprof format. Allowed values: [‘SHELX’, ‘Fullprof’]

OutputFile

Input

string

Filepath to save the integrated peaks workspace in HKL format

OutputWorkspace

Output

Workspace

Mandatory

Output peaks workspace (copy of input with updated peak intensities)

Description

Uses a MDEventWorkspace and PeaksWorkspace to integrate the peaks provided in the peaks workspace. A LorentzCorrection can be optionally applied to the integrated peaks with the ApplyLorentz option.

The output peaks workspace can be written to OutputFile in either SHELX format with direction cosines using SaveHKL or in the Fullprof format using SaveReflections.

The input to this algorithm is intended as part of the DEMAND data reduction workflow, using HB3AAdjustSampleNorm and HB3AFindPeaks or HB3APredictPeaks which can be seen below in the example usage.

You can make use of ConvertQtoHKLMDHisto to view the data in HKL.

Workflow

There are a few ways to go about processing DEMAND (HB3A) data. First you can simply load the data with HB3AAdjustSampleNorm, then predict the peaks positions with HB3APredictPeaks using the existing UB in the data workspace, then integrate peaks with HB3AIntegratePeaks.

If you don’t have the correct UB you can use HB3AFindPeaks to find peaks and the UB, then integrate the peaks found with HB3AIntegratePeaks. Alternatively, once you found the new UB you can use it to predict all expected peaks with HB3APredictPeaks.

../_images/HB3AWorkflow.svg

Usage

Example - DEMAND Workflow predicting peaks

# Input detector scan data to use. Can be a list of files, or workspaces - see HB3AAdjustSampleNorm for details
data_files = "HB3A_exp0724_scan0182.nxs, HB3A_exp0724_scan0183.nxs"

# Converts to MDEventWorkspace
#  If needed, the sample position can be adjusted by a height and distance - see the HB3AAdjustSampleNorm docs
data = HB3AAdjustSampleNorm(Filename=data_files)

# Finds peaks on the combined workspace, optimizing the UB matrix for the given cell type parameters
#  lattice parameters can be optionally specified to calculate the UB matrix (see docs for HB3AFindPeaks)
peaks = HB3APredictPeaks(InputWorkspace=data,
                         ReflectionCondition='B-face centred')

# Integrate the predicted peaks from the workspace above
integrated_peaks = HB3AIntegratePeaks(InputWorkspace=data, PeaksWorkspace=peaks,
                                      PeakRadius=0.25,
                                      OutputFile="./integrated_peaks.hkl")

# To visualize the combined data sets you can merge them together and overlay the integrated peaks in sliceviewer
merged = MergeMD(data)
# You can also convert the merged data in HKL to visualize the HKL data and overlay the integrated peaks in sliceviewer, see figure below
HKL = ConvertQtoHKLMDHisto(merged,
                           Extents='-5.1,5.1,-2.1,2.1,-20.1,20.1',
                           Bins='251,101,1001')
../_images/HB3A_exp0724.png

Example - DEMAND Workflow predicting satellite peaks

Following on from the previous example we will predict some (fictional) satellite peaks.

# Using a modulation vector of ±0.4H.
satellites_peaks = HB3APredictPeaks(InputWorkspace=data,
                                    ReflectionCondition='B-face centred',
                                    SatellitePeaks=True,
                                    ModVector1='0.4,0,0',
                                    MaxOrder=1,
                                    IncludeIntegerHKL=False)
integrated_satellites_peaks = HB3AIntegratePeaks(InputWorkspace=data, PeaksWorkspace=satellites_peaks,
                                                 PeakRadius=0.15,
                                                 OutputFormat="Fullprof",
                                                 OutputFile="./integrated_satellite_peaks.hkl")

You will see the modulation vectors in the output file

head -15 ./integrated_satellite_peaks.hkl

gives

integrated_satellites_peaks
(4i4,2f12.2,i5,4f10.4)
  1.00800 0 0
       2
    1     0.400000     0.000000     0.000000
    2    -0.400000    -0.000000    -0.000000
 #  h   k   l   m      Fsqr       s(Fsqr)   Cod   Lambda
   -1   1 -11   1        0.56        1.00    1    1.0080
    0   1 -12   2        3.60        2.45    1    1.0080
    0   1 -10   2       10.75        4.58    1    1.0080
    0   1  -8   2        2.94        2.65    1    1.0080
    0   0 -12   1       17.45        5.48    1    1.0080
    0   0 -10   1       28.58        7.62    1    1.0080
    0   0  -8   1       43.19       10.39    1    1.0080
    0   1 -10   1       17.98        5.83    1    1.0080

Red is the predicted nuclear peaks and green is the predicted satellite peaks

../_images/HB3A_exp0724_satellites.png

Example - DEMAND Workflow finding peaks and determining the UB

# Input detector scan data to use. Can be a list of files, or workspaces - see HB3AAdjustSampleNorm for details
data_files = "HB3A_exp0724_scan0182.nxs, HB3A_exp0724_scan0183.nxs"

# Converts to MDEventWorkspace
#  If needed, the sample position can be adjusted by a height and distance - see the HB3AAdjustSampleNorm docs
data = HB3AAdjustSampleNorm(Filename=data_files)

# Finds peaks on the combined workspace, optimizing the UB matrix for the given cell type parameters
#  lattice parameters can be optionally specified to calculate the UB matrix (see docs for HB3AFindPeaks)
peaks = HB3AFindPeaks(InputWorkspace=data,
                      CellType="Orthorhombic",
                      Centering="F")

# Integrate the peaks from the optimized peaks workspace above
integrated_peaks = HB3AIntegratePeaks(InputWorkspace=data,
                                      PeaksWorkspace=peaks,
                                      PeakRadius=0.25,
                                      OutputFile="./integrated_peaks.hkl")

# Alternatively, predict the peaks using the newly found UB, then integrate
predicted_peaks = HB3APredictPeaks(InputWorkspace=data,
                                   UBWorkspace=peaks,
                                   ReflectionCondition='B-face centred')
integrated_predicted_peaks = HB3AIntegratePeaks(InputWorkspace=data,
                                                PeaksWorkspace=predicted_peaks,
                                                PeakRadius=0.25,
                                                OutputFile="./integrated_peaks.hkl")

Categories: AlgorithmIndex | Crystal\Integration

Source