\(\renewcommand\AA{\unicode{x212B}}\)
Table of Contents
Name | Direction | Type | Default | Description |
---|---|---|---|---|
Filename | Input | string | The name of the input event Nexus file to load. Allowed extensions: [‘_event.nxs’] | |
InputWorkspace | Input | EventWorkspace | Input event workspace. Assumed to be unmodified events straight from LoadEventNexus | |
OutputWorkspace | Output | MatrixWorkspace | Mandatory | Then name of the output EventWorkspace |
NoBeamCenter | Input | boolean | False | If true, the detector will not be moved according to the beam center |
UseConfigBeam | Input | boolean | False | If true, the beam center defined in the configuration file will be used |
BeamCenterX | Input | number | Optional | Beam position in X pixel coordinates (used only if UseConfigBeam is false) |
BeamCenterY | Input | number | Optional | Beam position in Y pixel coordinates (used only if UseConfigBeam is false) |
UseConfigTOFCuts | Input | boolean | False | If true, the edges of the TOF distribution will be cut according to the configuration file |
LowTOFCut | Input | number | 0 | TOF value below which events will not be loaded into the workspace at load-time |
HighTOFCut | Input | number | 0 | TOF value above which events will not be loaded into the workspace at load-time |
SkipTOFCorrection | Input | boolean | False | IF true, the EQSANS TOF correction will be skipped |
WavelengthStep | Input | number | 0.1 | Wavelength steps to be used when rebinning the data before performing the reduction |
UseConfigMask | Input | boolean | False | If true, the masking information found in the configuration file will be used |
UseConfig | Input | boolean | True | If true, the best configuration file found will be used |
CorrectForFlightPath | Input | boolean | False | If true, the TOF will be modified for the true flight path from the sample to the detector pixel |
PreserveEvents | Input | boolean | True | If true, the output workspace will be an event workspace |
SampleDetectorDistance | Input | number | Optional | Sample to detector distance to use (overrides meta data), in mm |
SampleDetectorDistanceOffset | Input | number | Optional | Offset to the sample to detector distance (use only when using the distance found in the meta data), in mm |
SampleOffset | Input | number | Optional | Offset to be applied to the sample position (use only when using the detector distance found in the meta data), in mm |
DetectorOffset | Input | number | Optional | Offset to be applied to the detector position (use only when using the distance found in the meta data), in mm |
LoadMonitors | Input | boolean | True | If true, the monitor workspace will be loaded |
LoadNexusInstrumentXML | Input | boolean | True | Reads the embedded Instrument XML from the NeXus file (optional, default True). |
OutputMessage | Output | string | ||
ReductionProperties | Input | string | __sans_reduction_properties |
Workflow algorithm that loads EQSANS event data and applies basic corrections to the workspace. The workflow proceeds as follows:
This algorithm is rarely called directly. It is called by SANSReduction.
Categories: AlgorithmIndex | Workflow\SANS