\(\renewcommand\AA{\unicode{x212B}}\)

DNSFlippingRatioCorr v1

../_images/ImageNotFound.png

Enable screenshots using DOCS_SCREENSHOTS in CMake

Summary

Performs flipping ratio correction on a given dataset using the NiCr data.

See Also

DNSComputeDetEffCorrCoefs

Properties

Name

Direction

Type

Default

Description

SFDataWorkspace

Input

MatrixWorkspace

Mandatory

A workspace with spin-flip experimental data from sample.

NSFDataWorkspace

Input

MatrixWorkspace

Mandatory

A workspace with non spin-flip experimental data from sample.

SFNiCrWorkspace

Input

MatrixWorkspace

Mandatory

A workspace with spin-flip NiCr data.

NSFNiCrWorkspace

Input

MatrixWorkspace

Mandatory

A workspace with non spin-flip NiCr data.

SFBkgrWorkspace

Input

MatrixWorkspace

Mandatory

A workspace with spin-flip background for NiCr.

NSFBkgrWorkspace

Input

MatrixWorkspace

Mandatory

A workspace with non spin-flip background for NiCr.

SFOutputWorkspace

Output

MatrixWorkspace

Mandatory

A workspace to save the corrected spin-flip data.

NSFOutputWorkspace

Output

MatrixWorkspace

Mandatory

A workspace to save the corrected non spin-flip data.

Description

Warning

This algorithm is being developed for a specific instrument. It might get changed or even removed without a notification, should instrument scientists decide to do so.

This algorithm applies flipping ratio correction to a given data workspaces. As a result, following workspaces will be created:

  • output workspace with corrected spin-flip data. Sample logs will be copied from the data spin-flip workspace.

  • output workspace with corrected non-spin-flip data. Sample logs will be copied from the data non-spin-flip workspace.

Flipping ratio correction is performed using the measurements of \(Ni_{0.89}\,Cr_{0.11}\) standard sample (hereafter NiCr). Background for NiCr must be also measured and provided to the algorithm as an input. Both, spin-flip anf non-spin-flip measurements are required. This algorithm performs the flipping ratio correction according to J. Appl. Cryst. 42, 69-84, 2009. Calculations are made in a following steps:

  1. Subtract Background from NiCr:

    \(N^{SF,\,NSF}_i = (N^{SF,\,NSF}_i)_{raw} - (B^{SF,\,NSF}_i)\)

    The Minus v1 algorithm is used for this step. In the case of negative result, the error message will be produced and the algorithm terminates.

  2. Calculate the correction coefficients:

    \(k_i = \frac{N^{NSF}_i}{N^{SF}_i} - 1\)

    The Divide v1 algorithm is used for this step.

  3. Apply correction to the data:

    \((I^{NSF}_i)_{corr} = I^{NSF}_i + \frac{1}{k_i}\cdot\left(I^{NSF}_i - I^{SF}_i\right)\)

    \((I^{SF}_i)_{corr} = I^{SF}_i - \frac{1}{k_i}\cdot\left(I^{NSF}_i - I^{SF}_i\right)\)

    where \(I^{SF,\,NSF}_i\) are the neutron counts in the SFDataWorkspace and NSFDataWorkspace, respectively.

Valid input workspaces

The input workspaces have to have the following in order to be valid inputs for this algorithm.

  • The same number of dimensions

  • The same number of spectra

  • The same number of bins

  • All given workspaces must have the same polarisation (algorithm checks for the ‘polarisation’ and ‘polarisation_comment’ sample logs)

  • All given workspaces must be normalized either to monitor counts or to experiment duration

  • All given workspaces must have the same kind of normalization (algorithm checks for ‘normalized’ sample log)

  • All given workspaces must have the appropriate flipper status (algorithm checks for ‘flipper’ sample log): spin-flip workspaces must have flipper ‘ON’ and non-spin-flip workspaces must have flipper ‘OFF’

If any of these conditions is not fulfilled, the algorithm terminates.

For the physically meaningful correction it is also important that these workspaces have the same slits size, detector bank rotation angle and the neutron wavelength. If some of these parameters are different, algorithm produces warning. If these properties are not specified in the workspace sample logs, no comparison is performed.

Usage

Example - Apply flipping ratio correction to a Vanadium run:

from os.path import join
import numpy as np

datapath = "/path/to/data/dns/rc36b_standard_dz"
coilcurrents = join(datapath, 'currents.txt')

# define input files.
sf_vanafile = join(datapath, 'dz29100525vana.d_dat')
nsf_vanafile = join(datapath, 'dz29100526vana.d_dat')

sf_bkgrfile = join(datapath, 'dz29100645leer.d_dat')
nsf_bkgrfile = join(datapath, 'dz29100646leer.d_dat')

sf_nicrfile = join(datapath, 'dz29100585nicr.d_dat')
nsf_nicrfile = join(datapath, 'dz29100586nicr.d_dat')

# load files to workspaces
sf_vana = LoadDNSLegacy(sf_vanafile, Normalization='duration', CoilCurrentsTable=coilcurrents)
nsf_vana = LoadDNSLegacy(nsf_vanafile, Normalization='duration', CoilCurrentsTable=coilcurrents)

sf_nicr = LoadDNSLegacy(sf_nicrfile, Normalization='duration', CoilCurrentsTable=coilcurrents)
nsf_nicr = LoadDNSLegacy(nsf_nicrfile, Normalization='duration', CoilCurrentsTable=coilcurrents)

sf_bkgr = LoadDNSLegacy(sf_bkgrfile, Normalization='duration', CoilCurrentsTable=coilcurrents)
nsf_bkgr = LoadDNSLegacy(nsf_bkgrfile, Normalization='duration', CoilCurrentsTable=coilcurrents)

# for a physically meaningful correction, we must subtract background from Vanadium
# this step is usually not required for other kinds of samples
sf_vana_bg = sf_vana - sf_bkgr
nsf_vana_bg = nsf_vana - nsf_bkgr

# apply correction
DNSFlippingRatioCorr(sf_vana_bg, nsf_vana_bg, sf_nicr, nsf_nicr, sf_bkgr, nsf_bkgr,
                     SFOutputWorkspace='sf_corrected', NSFOutputWorkspace='nsf_corrected')

# retrieve output workspaces
sf_corrected = mtd['sf_corrected']
nsf_corrected = mtd['nsf_corrected']

# calculate ratio of spin-flip to non-spin-flip
vana_ratio = sf_corrected/nsf_corrected

# ratio must be around 2, print first 5 points of the data array
print(np.around(vana_ratio.extractY()[:5]))

Output:

[[ 2.] [ 2.] [ 2.] [ 2.] [ 2.]]

Categories: AlgorithmIndex | Workflow\MLZ\DNS | CorrectionFunctions\SpecialCorrections

Source

Python: DNSFlippingRatioCorr.py